Air/Oxygen Blender

Features

• Suitable for use in Emergency Room and ICU where accurate oxygen concentration is required
• Double-ended proportioning valve enables easy adjustment of oxygen concentration from 21% to 100%.
• Two-stage pressure-balancing mechanism is adopted to deliver steady oxygen concentration.
• Two outlet ports and bleed valve allow a wide range of applications
• Internal bypass and alarm mechanism ensure safe use.
• Accuracy: ± 3% FS

Model NumberWorking Pressure RangeTotal Delivery Flow RangeOutlet ConnectionRemarks
GMX030E-AIR/O260 ~ 75 psi3 ~ 30 LPMCGA 1240 DISSISO standard
GMX120E-AIR/O260 ~ 75 psi5 ~ 120 LPMCGA 1240 DISSISO standard
GMX030U-AIR/O250 ~ 65 psi3 ~ 30 LPMCGA 1240 DISSUSA standard
GMX120U-AIR/O250 ~ 65 psi5 ~ 120 LPMCGA 1240 DISSUSA standard